groups:mg:temperature_controller [2018/05/22 15:12] – [Table] rholstgroups:mg:temperature_controller [2018/06/01 14:32] (current) – [In loop stability] rholst
Line 26: Line 26:
  
 The output of the Meerstetter is filtered as shown [[groups:mg:meerstetter#output_filter|here]]. The output of the Meerstetter is filtered as shown [[groups:mg:meerstetter#output_filter|here]].
 +
 +For the configuration file of the Meerstetter refer to ''\\AFS\iqo.uni-hannover.de\projects\magnesium\Projekte\PTB Ultrastable laser\Laser at IQO\meerstetter rack einschub''
 +
 +==== In loop stability ====
 +
 +The Meerstetter stabilises the 1560 nm rio planex laser diode running at 123.5mA to the target temperature with a standard deviation of ca. 0.23 mK.
 +