Cavity spectrum: plan/curved mirror

Caculates the cavity spectrum and the frequency distance between TEM_00 and higher order modes m+n. Two curved mirrors possible. Programm made by T. Legero.

Parameter

Step 1: Calculate the Free-Spectral-Rang and Gouy-Phase

$$FSR = \frac{c}{2\cdot L}$$ $$Gouy = \frac{FSR}{\pi} \cdot \arccos \left( \sqrt{(1-\frac{L}{R1})\cdot(1-\frac{L}{R2}} \right)$$

Step 2: Calculate the frequency difference between TEM_00 and TEM_mn

$$ \Delta = FSR - mn\cdot Gouy $$

$$ \Delta(mn,w) = w\cdot FSR - mn\cdot Gouy $$

Freq. diff. of m+n = 4 higher order mode to 00-mode is 7.95436 MHz
Freq. diff. of m+n = 37 higher order mode to 00-mode is -4.49309 MHz
Freq. diff. of m+n = 41 higher order mode to 00-mode is 3.46128 MHz
Freq. diff. of m+n = 74 higher order mode to 00-mode is -8.98616 MHz
Freq. diff. of m+n = 78 higher order mode to 00-mode is -1.03181 MHz
Freq. diff. of m+n = 82 higher order mode to 00-mode is 6.92256 MHz

Step 3: Calculate the frequency difference between TEM_00 and the nearest TEM_nm

$$ \Delta = FSR - mn\cdot Gouy $$

Achtung: ROC wird hoffentlich >1m werden, damit die m+n = 4 Mode weiter weg ist!