The optical amplifiers are limited by non-linear behavior above a voltage of 3V
For the optical amplifier “ADA4898-1” at U2 following resistors and capacitors have to be placed, so that 3 V aren't overridden:
C2: 1 kΩ, C3: 100 pF, R4: 1 kΩ, R8: 0 Ω, C20: 0 Ω, R5: 100 Ω
For the optical amplifier “OPA656” at U1 the resistor R11 has to be changed to 200 kΩ
instead of the optical amplifier “ADA4898-1” the “OP27” is used at the position U3
For the optical amplifier “OP27” at U3 only slot 1 at the dipswitch has to be turned on.
During the setup of the photodiode, this dipswitch was changed, as the capacity of the incoming light was lower than expected. So now slot 2 is turned on instead of 1, so that the amplification factor is 10 instead of 1. (info: not shown in the picture below)