====== Reflectiondip / Impedance of the cavity ======
==== Annahme: ====
==== Parameter: ====
^ Bezeichnung ^ Variable ^
| Finesse | F |
| Transmission | T |
| Absorption | A |
| Reflektion | R |
| Diptiefe | DIP |
| Absorptions-Intensität | IR |
| Einstrahl-Intensität | Iein |
| Transmissions-Intensität (Peakhöhe) | IT |
==== Werte: ====
^ Wert i ^ Ai [ppm] ^ Ti [ppm] ^ Comment zu Ai und Ti ^ F ^ Comment zu F ^
| 1 | 18 | 8 | Werte werden angenommen. | - | Ist zu berechnen. |
| 2 | 20 | 8 | ::: | ::: | ::: |
| 3 | - | 11 | Transmisson abgelesen für Frequenz 1544nm aus: [[https://iqwiki.iqo.uni-hannover.de/lib/exe/detail.php?id=groups%3Amg%3Aproject_ptb-cavity%3Amirrors&media=groups:mg:project_ptb-cavity:1550_nm_epi_props.jpg|Transmission of -Polarized and p-polarized light at normal incidence]] | 107.000 | Aus: [[https://iqwiki.iqo.uni-hannover.de/doku.php?id=groups:mg:project_ptb-cavity:finesse_assessment_for_vacuum|Finesse assessment for vacuum]] |
| 4 | - | 11 | ::: | ::: | ::: |
==== Zusammenhänge: ====
$$ R = 1-T-A \;\text{Annahme: }\; R_{1}=R_{2}=R $$
$$ F = \pi \cdot \frac{\sqrt[4]{R_{1}\cdot R_{2}}}{1-\sqrt{R_{1}\cdot R_{2}}} -> F = \pi \cdot \frac{\sqrt{R}}{1-R}$$
Diptiefe aus dem Meschede: $$D_{IP}= 1-\frac{I_{R}}{I_{ein}}=1-\left( \frac{T_{1}+A_{1}+A_{2}-T_{2}}{T_{1}+A_{1}+A_{2}+T_{2}} \right)^{2}$$
$$ I_{T}=\frac{4\cdot T_{1}\cdot T_{2}\cdot \left( 1-A_{1}-A_{2} \right) }{\left( T_{1}+A_{1}+A_{2}+T_{2} \right)^{2}} $$
==== Berechnung der Finesse, Diptiefe und ... angenommen Werten 1 und 2: ====
Ti und Ai einsetzten und berechnen:
$$ D_{IP}=50\% $$
$$ I_{T}=9\% $$
$$ F=116.000 $$
Ergebnis: Ist ok, aber nicht ideal!
==== Berechnung der Absorbtion mit abgelesenen Werten 3 und 4: ====
Einsetzen von Finesse und Ti und berechnen:
$$ A_{3}=12ppm $$
$$ A_{4}=20ppm $$
$$ D_{IP}=60\% $$
$$ I_{T}=13\% $$