====== 02.11.2016 ====== * 8x sinkings for the vitonballs in each panel: {{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1351.jpg?300|}} * We measured the diameter of the glasballs and saw an inaccuracy of **80mm +/- 20müm**. Therefore we put the balls in a 5.Xmm test-hole and measured the ball-high in different position with the CNC-maschine: {{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1353.jpg?300|}} * After the "ball-high-measurement", we calculated the mean value and drilled again test-holes, for example "Kugel 1" **5.Xmm + 19.4müm**, and measured again. We saw a really nice result! After a 1cm test-milling and a good result, we milled in the mounting-breadboard: {{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1335.jpg?300|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1337.jpg?300|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1338.jpg?300|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1354.jpg?300|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1355.jpg?300|}} * We build everthing together and measured the high of the mounting on different places with a provisional result of **+/- ~80müm**: {{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1341.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1340.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1342.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1343.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1344.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1345.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1346.jpg?200|}}{{:groups:mg:project_ptb-cavity:labbook:2016-11:img_1347.jpg?200|}}