====== Highfinesse Wavemeter WS/6-200 ====== {{ :groups:mg:wavemeterhead.jpg?125|}} Installation: - Install with Compatibility mode for Windows 7 - Connect the USB cable * Possible problems: - USB cable is too long or must be plugged in again... - Perform the calibration with the Wutz laser (blue: 655.659 85 THz) in the AtomLab. The software is already installed on the Wavemeter computer. Otherwise you can find it in the AFS. * Possible problems: - You used the IR light from Wutz. The WS-6 can only be calibrated up to 900nm - The calibration is so long ago and the wavemeter is already too far away from the calibration frequency that one has to calibrate in small steps (100-50 MHz) from the current frequency to the nominal frequency. - Copy the entire folder Wavelength Meter WS 6-454 to ResLab computer - Start the program wlm_ws6.exe Calibration with the Wutz-light is necessary at least once a year! Otherwise the wavemeter doesn´t show the right frequency, because it has a drift! Accuracy: 200 MHz! Don´t forget to push the “Start” button {{:groups:mg:wavemeter.jpg|}}