====== 20-Channel-Multimeter/Rasmus-Box ====== ===== 20-Channel-Multimeter ===== * Company: Agilent * Main-device: 34972A * 20-Channel-Multimeter-Typ: 34970A (Multipexer) * [[https://www.keysight.com/en/pd-1000000083%3Aepsg%3Apro-pn-34901A/20-channel-multiplexer-2-4-wire-module-for-34970a-34972a?cc=DE&lc=ger|34970A]] * {{ :groups:mg:5965-5290dee.pdf |Product info_34970A}} * Typ: 2-polige Armatur (4-polig konfigurierbar) Zum Messverfahren bitte Seite 17 und 21 im Product info_34970A lesen! ^ Geschwindigkeit (Channels/s) ^ Max. Spannung ^ Max. Strom ^ Bandbreite ^ Thermospannung ^ Anmerkung ^ | 60/s | 300 V | 1 A | 10 MHz | <3 µV | Interner Thermo ele mentBezugspunkt \\ 2 zusätzliche Stromkanäle (insgesamt 22 Kanäle) | * For **NTC** and **Voltage** measurement use Channel **1-20** * For 4 wire measurent use Channel **1&10** or **11&20** * For **Current** measurement use Channel **21 and 22** (not connected in Rasmus-Box) ===== Rasmus-Box for NTC/Voltage measurements ===== Das Multimeter/Rasmus-Box wurde mit einem festen Widerstand (9,94 kOhm, vermessen mit BM805s Digitalmultimeter) getestet und die Channels wurden mit dem Programm //Agilent BVenchLink Data Logger3// vermessen. Das Ergbenis war, dass die Channels 1-10 [9942,68-9942,74 Ohm] einen 0,05-0,1 Ohm kleinen Widerstand angezeigt haben als die Channels 11-20 [9942,80 - 9942,85 Ohm]. Der Grund ist, dass die beiden Kabel zwischen Rasmus-Box und Multimeter vermutlich unterschiedlich lang sind. Alle Werte haben um circa 0,02 Ohm geschwankt!\\ {{:groups:mg:img_4659.jpg?300|}}{{:groups:mg:img_4660.jpg?300|}}{{:groups:mg:img_4661.jpg?300|}}\\ ==== Verbindungen ==== ^ Kabel zum Multimeter ^ Farbe ^ Belegung ^ Kanal ^ BNC ^ Anschlusskabel Box ^ Anschluss im Multimeter ^ | 1 | Schwarz | + | 01 | Pin | Weiss | H | | ::: | Violett | - | ::: | Shielding | Braun | L | | ::: | Grau-Braun | + | 02 | Pin | Weiss | H | | ::: | Weiß-Pink | - | ::: | Shielding | Braun | L | | ::: | Grau-Pink | + | 03 | Pin | Weiss | H | | ::: | Braun-Pink | - | ::: | Shielding | Braun | L | | ::: | Rot-Blau | + | 04 | Pin | Weiss | H | | ::: | Weiß | - | ::: | Shielding | Braun | L | | ::: | Weiß-Grau | + | 05 | Pin | Weiss | H | | ::: | Weiß-Grün | - | ::: | Shielding | Braun | L | | ::: | Pink | + | 06 | Pin | Weiss | H | | ::: | Rot | - | ::: | Shielding | Braun | L | | ::: | Blau | + | 07 | Pin | Weiss | H | | ::: | Gelb | - | ::: | Shielding | Braun | L | | ::: | Braun | + | 08 | Pin | Weiss | H | | ::: | Braun-Gelb | - | ::: | Shielding | Braun | L | | ::: | Braun-Grün | + | 09 | Pin | Weiss | H | | ::: | Weiß-Gelb | - | ::: | Shielding | Braun | L | | ::: | Grün | + | 10 | Pin | Weiss | H | | ::: | Grau | - | ::: | Shielding | Braun | L | | 2 | Schwarz | + | 11 | Pin | Weiss | H | | ::: | Violett | - | ::: | Shielding | Braun | L | | ::: | Grau-Braun | + | 12 | Pin | Weiss | H | | ::: | Weiß-Pink | - | ::: | Shielding | Braun | L | | ::: | Grau-Pink | + | 13 | Pin | Weiss | H | | ::: | Braun-Pink | - | ::: | Shielding | Braun | L | | ::: | Rot-Blau | + | 14 | Pin | Weiss | H | | ::: | Weiß | - | ::: | Shielding | Braun | L | | ::: | Weiß-Grau | + | 15 | Pin | Weiss | H | | ::: | Weiß-Grün | - | ::: | Shielding | Braun | L | | ::: | Pink | + | 16 | Pin | Weiss | H | | ::: | Rot | - | ::: | Shielding | Braun | L | | ::: | Blau | + | 17 | Pin | Weiss | H | | ::: | Gelb | - | ::: | Shielding | Braun | L | | ::: | Braun | + | 18 | Pin | Weiss | H | | ::: | Braun-Gelb | - | ::: | Shielding | Braun | L | | ::: | Braun-Grün | + | 19 | Pin | Weiss | H | | ::: | Weiß-Gelb | - | ::: | Shielding | Braun | L | | ::: | Grün | + | 20 | Pin | Weiss | H | | ::: | Grau | - | ::: | Shielding | Braun | L | ===== Zweite Box ===== Bei der zweiten Box wurde auf die Kabel zum anlöten von Temperatursensoren verzichtet. Die Charakterisierung der Box kann im AFS unter ''\\AFS\iqo.uni-hannover.de\projects\magnesium\Electronics\20 Kanal Multimeter'' gefunden werden. {{:groups:mg:private:resonatoren:multimeterbox:average_distribution_of_measured_values.png?400|}} **Kurzfassung:** Die Widerstandsmessung hat eine Standardabweichung von 0.0261 Ohm auf jedem Kanal. Die aus dem Datanblatt berechnete Genauigkeit ist 0.250 Ohm. Beim Messen von Widerständen müssen diese **Korrekturwerte** von den Messwerten abgezogen werden (in Ohm): 01: 0.09957333333659335\\ 02: 0.06472666666559235\\ 03: 0.07986166666887584\\ 04: 0.10997999999926833\\ 05: 0.13713833333349612\\ 06: 0.15628666666452773\\ 07: 0.15756166666506033\\ 08: 0.16503999999986263\\ 09: 0.1778100000010454\\ 10: 0.16759499999716354\\ 11: 0.36965333333318995\\ 12: 0.18343166666818433\\ 13: 0.3638133333333826\\ 14: 0.35798333333514165\\ 15: 0.3638133333333826\\ 16: 0.346864999999525\\ 17: 0.33629999999720894\\ 18: 0.3264516666677082\\ 19: 0.3341099999979633\\ 20: 0.3038266666681011 ==== Schirmung ==== Die Schirmung der Kabel muss in dem 20 Kanal Expander an die Masse wie folgt angeschlossen werden (grüne Litze): {{:groups:mg:private:resonatoren:multimeterbox:anschluss_schirmung_20ch_multimeter.jpg?400|}}