====== Checklist after power failure ====== If you know a power failure will come, than turn off everthing (without Meerstetter, IGP´s and H-Maser)! And organise a lot of USV´s!!!! **Check both IGP Controller for the resonators R1 + R2** * Current should be below 2 mA (both should show "HV ON" (green LED on) * Check both USV for IGP´S **Turn on the computers** **Check Meerstetter TEC for R1 + R2 + ECDL** * Green LED on? * Check in the software //TEC Service v2.70 - R1//, //TEC Service v2.70 - R2// and //TEC Service v2.70 - ECDL Baseplate// on the ResLab-PC: * How is the behavior of the temperature? * Check both USV for TEC´s **Check the air conditioner for the entire room** * Is the air-conditioning system switched on? * Is the V-belt (=Keilriemen) broken? * What is the temperature of the draft (=Zugluft)? -> should be >20 °C * Check the electric fuse (=Sicherung) on the third floor **Check the H-Maser** * All LED are green? * Check in the Software //VCH// on the ResLabMobil: * All values except for D2h must be green! **Check the 914nm-laser** * Light is on? Powermodule is on? * TA stabilization is ok? * Laserdiode is on? * How much power is in front of the resonator? -> should be 70-100 µW * What wavelength does the Wavemeter display? **Check the 10 MHz reference and set all frequencies in the lab** * GPS-Receiver is working? * LED flashes orange/yellow? * If not Turn on the PWR switch until LED flashes * Start the software //Net View// and connect the receiver (COM Port2) * Rb-clock is working? * Check the LED´s: * //Power//-LED and //Locked//-LED have to be green * //1 pps sync//-LED has to be green * //1 pps input//-LED has to flash * Check the 10 MHz-frequency output with an spectrum analyzer, if there is 10 MHz. * Check the 10 MHz distributor synthesizer * Ext refernce //10 MHz direct// is working? * Set correct values * Check all other synthesizer * Ext refernce //10 MHz direct// is working? * Set correct values **Check frequency comb** * Laser and amplifier are on? * Check LED´s * Do the software //fiber laser// and //fiber comb control// work? * Can you speak with the comb? **Check the counter** * Does the software //FXQE80// work? * If not: Turn off and on the counter (cold-device plug disconnect and connect) **Check the fiber Link to PTB** * Turn everthing on, like ther spectrum analysizer * Restart the PC in the Atom Lab * Call them! **If the temperature is normal again, then...** - Stabilize the laser of R1 - Turn on the Laserdiode - Turn on the TA - Couple the light into the TA - Turn on the TA stabilization - Lock R1 - Stablize the frequency comb - Lock Ceo - Lock Rep - Measure the frequencies with the counter: - 914nm - frep - 10 MHz - virtuel beat