====== Aktive vibration isolation ====== Max 100kg! We have two stages: R1 & clock laser are on one stage and R2 is on another stage. It is a special design because the electronics have been removed of the vibration platform for thermal reasons.\\ {{:groups:mg:img_1866.jpg?400|}} ===== Modi ===== * three modi available: **aktive**, **transportable**, **off** * Information on **aktive** modus: - First step: Complete load adjust with "up"-button * load adjust: //on// or //off// or //locked// - Second step: Activate isolation with "down"-button * isolation //on// or //off// Important in **aktive modus**: The "four bars" VS, HS, VA, HA should be in the upper third of the range for an adequate control range. * Information on **transportable** modus: Mechanic is decoupled. Now is working, for example optics screw down, possible. * Information on **off** modus: - Movements in the room have influence of the vibration isolation! ===== Manual and datasheet ===== * {{ :groups:mg:private:resonatoren:aktiv_vibration_isolation:flyer_halcyonics_micro_series_1_.pdf | Flyer}} * {{ :groups:mg:private:resonatoren:aktiv_vibration_isolation:tech_spec_halcyonics_micro_1_.pdf | Datasheet}} * {{ :groups:mg:private:resonatoren:aktiv_vibration_isolation:operation_manual_halcyonics_micro_series_1_.pdf | Manual}}