groups:mg:project_ptb-cavity:labbook:2017-02:labbookentry-2017-02-10 [2017/02/10 12:51] pkoopmanngroups:mg:project_ptb-cavity:labbook:2017-02:labbookentry-2017-02-10 [2017/02/10 13:38] (current) – [10.02.2017 - Auswertung der Lasertreibercharakterisierung] pkoopmann
Line 8: Line 8:
   * Fluke Multimeter 112: 1% +/- 0,5 mA   * Fluke Multimeter 112: 1% +/- 0,5 mA
  
-Gleichung für den Knopf: Y = -0.118 +/- 0,44 * X \\ +Gleichung für den Knopf in Abhängigkeit des Laserstroms: Y = -0.118 + 0,044 * X \\ 
-Gleichung für den Display: Y = -1,366 +/- 1,004 * X \\ +Gleichung für den Display in Abhängigkeit des Laserstroms: Y = -1,366 + 1,004 * X \\
 {{:groups:mg:project_ptb-cavity:labbook:2017-02:lasertreiberauswertung.jpg?600|}} {{:groups:mg:project_ptb-cavity:labbook:2017-02:lasertreiberauswertung.jpg?600|}}
 +\\
 +**Auswertungen:**
 +  * Obwohl der Laserdiodenstrom auf 200 mA begrenzt ist, wurden mehr als 200 mA gemessen: maximal gemessener Strom: 228 mA +/- 2,78
 +  * Der Laserdiodenstrom ist höher, als der Display des Lasers anzeigt. \\ Berechnung der Abweichung des Laserstroms: X= (Y + 1,366)/1,004 (Y ist der Wert des Displays)
 +
 +**Nächste Schritte:**
 +  * Klären, warum der Laserdiodenstrom beim Anschalten auf das Maximum springt
 +  * Maximum des Laserstroms einstellen