Line 1: Line 1:
 ====== Highfinesse Wavemeter WS/6-200 ======  ====== Highfinesse Wavemeter WS/6-200 ====== 
 {{  :groups:mg:wavemeterhead.jpg?125|}} {{  :groups:mg:wavemeterhead.jpg?125|}}
 +Installation:
 +  - Install with Compatibility mode for Windows 7
 +  - Connect the USB cable
 +    * Possible problems:
 +       - USB cable is too long or must be plugged in again... 
 +  - Perform the calibration with the Wutz laser (blue: 655.659 85 THz) in the AtomLab. The software is already installed on the Wavemeter computer. Otherwise you can find it in the AFS.
 +    * Possible problems:
 +       - You used the IR light from Wutz. The WS-6 can only be calibrated up to 900nm
 +       - The calibration is so long ago and the wavemeter is already too far away from the calibration frequency that one has to calibrate in small steps (100-50 MHz) from the current frequency to the nominal frequency.
 +   - Copy the entire folder Wavelength Meter WS 6-454 to ResLab computer
 +   - Start the program wlm_ws6.exe
 +
  
 <note important>Calibration with the Wutz-light is necessary at least once a year! Otherwise the wavemeter doesn´t show the right frequency, because it has a drift! </note>  <note important>Calibration with the Wutz-light is necessary at least once a year! Otherwise the wavemeter doesn´t show the right frequency, because it has a drift! </note> 
 +
 +
 +<note important>Accuracy: 200 MHz! </note> 
 <note important>Don´t forget to push the “Start” button</note>  <note important>Don´t forget to push the “Start” button</note> 
  
 {{:groups:mg:wavemeter.jpg|}} {{:groups:mg:wavemeter.jpg|}}