====== Highfinesse Wavemeter WS/6-200 ====== | ====== Highfinesse Wavemeter WS/6-200 ====== | ||
{{ : | {{ : | ||
+ | Installation: | ||
+ | - Install with Compatibility mode for Windows 7 | ||
+ | - Connect the USB cable | ||
+ | * Possible problems: | ||
+ | - USB cable is too long or must be plugged in again... | ||
+ | - Perform the calibration with the Wutz laser (blue: 655.659 85 THz) in the AtomLab. The software is already installed on the Wavemeter computer. Otherwise you can find it in the AFS. | ||
+ | * Possible problems: | ||
+ | - You used the IR light from Wutz. The WS-6 can only be calibrated up to 900nm | ||
+ | - The calibration is so long ago and the wavemeter is already too far away from the calibration frequency that one has to calibrate in small steps (100-50 MHz) from the current frequency to the nominal frequency. | ||
+ | - Copy the entire folder Wavelength Meter WS 6-454 to ResLab computer | ||
+ | - Start the program wlm_ws6.exe | ||
+ | |||