===== Whole IQ/UNI power failure ===== | ===== Whole IQ/UNI power failure ===== | ||
- | <note important> | + | <note important> |
* Turn off every device (Oscilloscopes, | * Turn off every device (Oscilloscopes, | ||
* Check if the AC is working again (call 4440 to tell them or wait for the next day until it's fixed).< | * Check if the AC is working again (call 4440 to tell them or wait for the next day until it's fixed).< | ||
Line 28: | Line 28: | ||
**Triplet Laser** | **Triplet Laser** | ||
* Turn on the Thorlabs TECs for the Laserdiodes (they turn off on a power failure) | * Turn on the Thorlabs TECs for the Laserdiodes (they turn off on a power failure) | ||
- | * Turn off (even when the controlers are still off) the laser-diodes | + | * Turn down the current for every laser-diodes |
+ | * Reenable the powersupply for the LD (in the Stromschrank): | ||
+ | * Turn on each laser controler. Then un-short the diodes. The ramp up the current | ||
+ | * If a PID does not seem to work (e.g. Offset cant be set): Unplug it from power and replug it. | ||
**Wutz** | **Wutz** | ||
* Turn on the Toptica controler (green button) - If it does turn off itself: Wait 15 minutes for the TECs to settle and try aganin. | * Turn on the Toptica controler (green button) - If it does turn off itself: Wait 15 minutes for the TECs to settle and try aganin. | ||
* Turn on the EOM-Synthesizer (located on bottom of Dipole-Rack) and set it to the correct frequency (written on the device - Recall, 9, Enter) | * Turn on the EOM-Synthesizer (located on bottom of Dipole-Rack) and set it to the correct frequency (written on the device - Recall, 9, Enter) | ||
+ | * Monitor for Wutz/ | ||
+ | **Lattice Laser** | ||
+ | * Check if the Meerstetter for the TiSa is working (green LEDs) | ||
+ | * If a PID does not seem to work (e.g. Offset cant be set): Unplug it from power and replug it. |