- | ====== At power failure ====== | + | ====== At (severe/ |
* Set up the emergency power system (EPS) on the balcony room - (Parameters to turn it on: Choke: Turn off, Gas: Turn on) | * Set up the emergency power system (EPS) on the balcony room - (Parameters to turn it on: Choke: Turn off, Gas: Turn on) | ||
* Disconnect the marked power-cord (next to " | * Disconnect the marked power-cord (next to " | ||
* If the turbo-pumps fail: Close the big valve below the chamber-table for the big turbo pump and turn off the big turbo. | * If the turbo-pumps fail: Close the big valve below the chamber-table for the big turbo pump and turn off the big turbo. | ||
- | ====== | + | ====== |
- | <note important> | + | There are two power-failure situations. One, where a fuse or the Notaus in the lab has triggered the power failure and then one, which happens for the whole IQ/UNI. The later one needs an additional treatment. |
+ | |||
+ | ===== Whole IQ/UNI power failure | ||
+ | <note important> | ||
* Turn off every device (Oscilloscopes, | * Turn off every device (Oscilloscopes, | ||
* Check if the AC is working again (call 4440 to tell them or wait for the next day until it's fixed).< | * Check if the AC is working again (call 4440 to tell them or wait for the next day until it's fixed).< | ||
+ | * If ACs are working again, turn on every single device again. | ||
+ | * | ||
+ | ===== Checklist of devices ===== | ||
+ | **Vacuum** | ||
+ | * Check if all pumps are running and all pressures are ok | ||
+ | * If pressures and pumps are ok (especially pre-vacuum and turbos) open the pneumatic valves (Next to the Lab-Door, there is a orange interlock. Press the reset button. You will hear the valves) | ||
+ | * Open, if it has been closed, the big valve below the table at the big turbopump | ||
+ | |||
+ | **S-MOT Laser** | ||
+ | * Turn on the Toptica controler (green button) | ||
+ | * Check the temperature controler of 285 | ||
+ | * If a PID does not seem to work (e.g. Offset cant be set): Unplug it from power and replug it. | ||
+ | |||
+ | **Triplet Laser** | ||
+ | * Turn on the Thorlabs TECs for the Laserdiodes (they turn off on a power failure) | ||
+ | * Turn down the current for every laser-diodes (even when the controlers are still off), turn the controlers off and set the short button to short | ||
+ | * Reenable the powersupply for the LD (in the Stromschrank): | ||
+ | * Turn on each laser controler. Then un-short the diodes. The ramp up the current | ||
+ | * If a PID does not seem to work (e.g. Offset cant be set): Unplug it from power and replug it. | ||
+ | |||
+ | **Wutz** | ||
+ | * Turn on the Toptica controler (green button) - If it does turn off itself: Wait 15 minutes for the TECs to settle and try aganin. | ||
+ | * Turn on the EOM-Synthesizer (located on bottom of Dipole-Rack) and set it to the correct frequency (written on the device - Recall, 9, Enter) | ||
+ | * Monitor for Wutz/ | ||
+ | |||
+ | **Lattice Laser** | ||
+ | * Check if the Meerstetter for the TiSa is working (green LEDs) | ||
+ | * If a PID does not seem to work (e.g. Offset cant be set): Unplug it from power and replug it. |