meta data for this page
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
groups:mg:htc-3000 [2020/02/05 12:18] – [Power Supply] fiene | groups:mg:htc-3000 [2024/03/20 09:36] (current) – Admin: Syntax-Update (Migration from deprecated "fontcolor" plugin to "color" plugin) klaus | ||
---|---|---|---|
Line 12: | Line 12: | ||
The right switch has two positions: | The right switch has two positions: | ||
- | * In the upper position the current voltage output by the HTC in order to monitor the current temperature is displayed. | + | * In the lower position the current voltage output by the HTC in order to monitor the current temperature is displayed. |
- | * In the lower position the voltage corresponding to the wanted temperature is displayed. | + | * In the upper position the voltage corresponding to the wanted temperature is displayed. |
- | While you <fc # | + | While you <color # |
The left switch enables or disables the output temperature. Turn off while changing the set temperature.\\ | The left switch enables or disables the output temperature. Turn off while changing the set temperature.\\ | ||
Line 76: | Line 76: | ||
The necessary temperatures are: | The necessary temperatures are: | ||
^ SHG ^ Temperature | ^ SHG ^ Temperature | ||
- | | #1 | 38,6 °C | 0,48 V | | + | | #1 |
- | | #2 | 29,3 °C | 0,97 V | | + | | #2 |
- | | #3 | 18,8 °C | | | + | | #3 |
(Status of 04.02.2020) | (Status of 04.02.2020) | ||
The temperaturecontrolling processes somewhat like this: \\ | The temperaturecontrolling processes somewhat like this: \\ | ||
- | {{: | + | {{ :groups:mg:temperaturecontrollingdiagramm.pdf |Characterization of SHG #3}} |
===== Replacing the HTC ===== | ===== Replacing the HTC ===== | ||
Unscrew the HTC and detach the pins. You may also have to unscrew the heatsink from the metal shroud. Then plug the new HTC back in and tighten all screws firmly. | Unscrew the HTC and detach the pins. You may also have to unscrew the heatsink from the metal shroud. Then plug the new HTC back in and tighten all screws firmly. | ||
Line 91: | Line 91: | ||
| **Problem** | **Possible Solution(s)** | | | **Problem** | **Possible Solution(s)** | | ||
- | | When viewing " | + | | When viewing " |
| Heating and cooling are mixed up | Change the polarity of the peltier element. See page 8 of the {{ : | | Heating and cooling are mixed up | Change the polarity of the peltier element. See page 8 of the {{ : | ||
| The Element neither heats nor cools | Check the connector, test the element. Otherwise see Power Supply Related Problems | | | The Element neither heats nor cools | Check the connector, test the element. Otherwise see Power Supply Related Problems | |