-====== Checklist after power failure ====== 
- 
-<note important>If you know a power failure will come, than turn off everthing (without Meerstetter, IGP´s and H-Maser)! And organise a lot of USV´s!!!!</note> 
- 
- 
-**Check both IGP Controller for the resonators R1 + R2** 
-    * Current should be below 2 mA (both should show "HV ON" (green LED on) 
-    * Check both USV for IGP´S 
- 
-**Turn on the computers** 
- 
-**Check Meerstetter TEC for R1 + R2 + ECDL**  
-    * Green LED on?  
-    * Check in the software //TEC Service v2.70 - R1//, //TEC Service v2.70 - R2// and //TEC Service v2.70 - ECDL Baseplate// on the ResLab-PC:  
-      * How is the behavior of the temperature? 
-    * Check both USV for TEC´s 
-  
-**Check the air conditioner for the entire room** 
-    * Is the air-conditioning system switched on? 
-      * Is the V-belt (=Keilriemen) broken? 
-    * What is the temperature of the draft (=Zugluft)? -> should be >20 °C 
-      * Check the electric fuse (=Sicherung) on the third floor 
- 
-**Check the H-Maser** 
-    * All LED are green? 
-    * Check in the Software //VCH// on the ResLabMobil: 
-      *  All values except for D2h must be green! 
- 
-**Check the 914nm-laser** 
-    * Light is on? Powermodule is on?   
-    * TA stabilization is ok? 
-    * Laserdiode is on? 
-    * How much power is in front of the resonator? -> should be 70-100 µW 
-    * What wavelength does the Wavemeter display? 
- 
-**Check the 10 MHz reference and set all frequencies in the lab** 
-    * GPS-Receiver is working? 
-      * LED flashes orange/yellow? 
-        * If not Turn on the PWR switch until LED flashes 
-        * Start the software //Net View// and connect the receiver (COM Port2) 
-    * Rb-clock is working? 
-      * Check the LED´s:  
-        * //Power//-LED and //Locked//-LED have to be green 
-        * //1 pps sync//-LED has to be green 
-        * //1 pps input//-LED has to flash 
-      * Check the 10 MHz-frequency output with an spectrum analyzer, if there is 10 MHz. 
-    * Check the 10 MHz distributor synthesizer 
-      * Ext refernce //10 MHz direct// is working? 
-      * Set correct values 
-    * Check all other synthesizer 
-      * Ext refernce //10 MHz direct// is working? 
-      * Set correct values   
- 
-**Check frequency comb** 
-     * Laser and amplifier are on? 
-       * Check LED´s 
-     * Do the software //fiber laser// and //fiber comb control// work? 
-       * Can you speak with the comb? 
- 
-**Check the counter** 
-  * Does the software //FXQE80// work? 
-   * If not: Turn off and on the counter (cold-device plug disconnect and connect) 
- 
-**Check the fiber Link to PTB** 
-  * Turn everthing on, like ther spectrum analysizer 
-  * Restart the PC in the Atom Lab 
-  * Call them! 
- 
-**If the temperature is normal again, then...** 
-  - Stabilize the laser of R1 
-    - Turn on the Laserdiode 
-    - Turn on the TA 
-    - Couple the light into the TA 
-    - Turn on the TA stabilization 
-    - Lock R1 
-  - Stablize the frequency comb 
-    - Lock Ceo 
-    - Lock Rep 
-  - Measure the frequencies with the counter: 
-    - 914nm 
-    - frep 
-    - 10 MHz 
-    - virtuel beat 
- 
- 
-