-====== Checklist after power failure ====== 
- 
-  - **Check both IGP Controller for the resonators R1 + R2** 
-    * Current should be below 2 mA (both should show "HV ON" (green LED on) 
-    * Check both USV for IGP´S 
- 
-  - **Check Meerstetter TEC for R1 + R2 + ECDL**  
-    * Green LED on?  
-    * Check in the software //TEC Service v2.70 - R1//, //TEC Service v2.70 - R2// and //TEC Service v2.70 - ECDL Baseplate// on the ResLab-PC:  
-      * How is the behavior of the temperature? 
-    * Check both USV for TEC´s 
-  
-  - **Check the air conditioner for the entire room** 
-    * Is the air-conditioning system switched on? 
-      * Is the V-belt (=Keilriemen) broken? 
-    * What is the temperature of the draft (=Zugluft)? -> should be >10 °C 
-      * Check the electric fuse (=Sicherung) on the third floor 
- 
-  - **Check the H-Maser** 
-    * All LED are green? 
-    * Check in the Software //VCH// on the ResLabMobil: 
-      *  All values except for D2h must be green! 
- 
-  - **Check the 914nm-laser** 
-    * Light is on? Powermodule is on?   
-    * TA stabilization is ok? 
-    * Laserdiode is on? 
-    * How much power is in front of the resonator? -> should be 70-100 µW 
-    * What wavelength does the Wavemeter display? 
- 
-  - Check the 10 MHz reference 
-    *