groups:mg:experiment:vacuum [2018/08/23 17:28] – njha | groups:mg:experiment:vacuum [2018/11/01 10:25] (current) – [Oil change of the pre-vacuum pump] fim | ||
---|---|---|---|
Line 52: | Line 52: | ||
* Now fill the oven to the top and again fix back the nut with the nozzle on it. | * Now fill the oven to the top and again fix back the nut with the nozzle on it. | ||
* Reconnect the oven flange to the vacuum chamber with a fresh copper gasket. Make sure that the vent gas is turned off before fixing the screws to avoid overpressure in the chamber. | * Reconnect the oven flange to the vacuum chamber with a fresh copper gasket. Make sure that the vent gas is turned off before fixing the screws to avoid overpressure in the chamber. | ||
+ | |||
+ | {{: | ||
+ | |||
===== Heating the pumps (usefull for low 10^-9 pressures and below and after opening the vacuum) ===== | ===== Heating the pumps (usefull for low 10^-9 pressures and below and after opening the vacuum) ===== | ||
Line 145: | Line 148: | ||
*Turbomolekularpumpe TMU 260: {{: | *Turbomolekularpumpe TMU 260: {{: | ||
#*In Betrieb seit XY | #*In Betrieb seit XY | ||
+ | |||
+ | |||
+ | |||
+ | ====== Oil change of the pre-vacuum pump ====== | ||
+ | |||
+ | The oil has to be changed **every third year** approximately. | ||
+ | |||
+ | - First, close the big yellow vent at the source chamber. Second, close the pneumatik vents via the knob at the orange power outlet (close to the door) and check if both are really closed. | ||
+ | - CLose the black vent of the pump which is placed close to the on/off switch. | ||
+ | - Open the drain screw (No. 8 in manul) labed as " | ||
+ | - Open the vents once the preassure at the pump is the same as at the source chamber. | ||
+ | |||
+ | Manual: {{ : | ||
+ | |||
+ | Last changes: nov 2018 | ||
+ |