meta data for this page
Differences
This shows you the differences between two versions of the page.
Previous revision | |||
— | groups:mg:experiment:laser:383nmlaser [2022/04/01 09:07] (current) – fim | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | ====== 383 nm Laser system (T-MOT) ====== | ||
+ | |||
+ | The 383 nm laser system consists of 767 nm lasers which are then frequency doubled to achieve 383 nm light. | ||
+ | |||
+ | **External Cavity Diode Laser (ECDL)** | ||
+ | |||
+ | For 767 nm lasers, ECDL in Littrow configuration is used. Typically we used the laser diodes from Eagleyard Photonics: EYP-RWE-0790-02000-1500-SOT02-0000 | ||
+ | |||
+ | Recently, Eagleyard has replaced these with new laser diodes: EYP-RWE-0760-02010-1500-SOT12-0000 | ||
+ | |||
+ | **TA** | ||
+ | * Output Power: 1.5 W | ||
+ | * Input Current: 2 A | ||
+ | * Injection Power: 32 mW | ||
+ | * Power behind 30dB Isolator: 1.05 W | ||
+ | * Originally this TA was used: EYP-TPA-0765-01500-3006-CMT03-0000. Is this still true? | ||
+ | |||
+ | **Fiber** | ||
+ | * PMC-780-5, | ||
+ | * Incoupling: 67% | ||
+ | * Power behind fiber: 700 mW | ||
+ | |||
+ | ===== Frequenzy doubling ===== | ||
+ | |||
+ | **LBO-Crystal** | ||
+ | * Lenght: 15 mm | ||
+ | * AR coating | ||
+ | |||
+ | **Resonator** | ||
+ | * Ring resonator (double Z configuration) | ||
+ | * Length: 280mm | ||
+ | * Curvature of mirrors: 50 mm (S3 and S4) | ||
+ | * Distance of mirrors: 64 mm | ||
+ | * Waist: 30µm (crystal), 130µm (long arm) | ||
+ | * Transmission: | ||
+ | * Conversion efficiency ENL = 6.1*10^(-5)/ | ||
+ | $$E_{NL} = \kappa L_c k_1 h_m (B,\xi) $$ | ||
+ | * Linear losses: eL = 0.85(0.15) % | ||
+ | * Finesse: F = 270 | ||
+ | |||
+ | {{ : | ||
+ | |||
+ | ===== Stabilisation ===== | ||
+ | * PDH-Method | ||
+ | * Error signal at about 20 MHz | ||
+ | |||
+ | |||
+ | {{ : | ||
+ | |||
+ | ===== Stabilisation: | ||
+ | {{ : | ||
+ | Dopplerfree Saturation Spectroscopy on D1 line of Potassium: | ||
+ | |||
+ | * 1st the two beams with similar intensity are generated by the beam sampler and are send trough the glas cell | ||
+ | * one part of the initial beam is going trough the sampler to a double pass aom and then in the other direction trough the glas cell, crossing only one beam. Important: the beam coming from the aom must have a much higher intensity! | ||
+ | * the first two beams, coming from the sampler trough the cell are than monitored by the PD. The signal is substracted | ||
+ | * in or case e use the aom to create the peak on one side of the mainpeak. This is important for the lock-in scheme | ||
+ | |||
+ | {{ : | ||
+ | |||
+ | |||