groups:mg:experiment:laser:285nmlaser [2017/05/03 11:39] – [Turning on the Laser] wfriesengroups:mg:experiment:laser:285nmlaser [2022/04/06 05:51] (current) – [Locking to Iodine-Spectroscopy] wfriesen
Line 30: Line 30:
             * Move the crosshair to the second slope, rightclick on press //PID: Lock to Slop//.             * Move the crosshair to the second slope, rightclick on press //PID: Lock to Slop//.
  
 +
 +        * **Troubleshooting** 
 +            * If signal in DigiLock-Software does not update/ signal is frozen  -> click the sampling checkbar to activate
 +            * If signal in DigiLock-Software does not show the usual error-signal at correct frequency, but is saturated at left and right edge of the frame (see picture) -> Check the ref input for the lock-in-amplifier (near the iodine spectroscopy). Probably the input signal is too low. It should be slightly higher than 2 V (~2.2V works fine).{{ :groups:mg:experiment:laser:img-20210119-wa0001.jpg?400|}}
 ===== Lasersystem - Typical power values ===== ===== Lasersystem - Typical power values =====
 The 285nm SHG was build by Jan Friebe and is described in his diplomathesis. The 285nm SHG was build by Jan Friebe and is described in his diplomathesis.