groups:mg:experiment:laser:285nmlaser [2017/01/05 11:51] – [285 nm-Laser (S-MOT)] fim | groups:mg:experiment:laser:285nmlaser [2022/04/06 05:51] (current) – [Locking to Iodine-Spectroscopy] wfriesen | ||
---|---|---|---|
Line 5: | Line 5: | ||
===== Turning on the Laser ===== | ===== Turning on the Laser ===== | ||
- | * Enable oxygen to 285nm SHG. | ||
* Enable main-switch. Now set the key to standby. | * Enable main-switch. Now set the key to standby. | ||
* Wait for the LED to change the color from orange to green to orange (~15 seconds after turning on). Turn the key now. | * Wait for the LED to change the color from orange to green to orange (~15 seconds after turning on). Turn the key now. | ||
Line 14: | Line 13: | ||
* //2 - 2.500 W// for //100-150 mW// directly after 285 nm SHG. | * //2 - 2.500 W// for //100-150 mW// directly after 285 nm SHG. | ||
* Do not exceed much more than //3.5 W// | * Do not exceed much more than //3.5 W// | ||
+ | * Flush the SHG with oxygen, by opening the bypass for roughly 10 seconds. | ||
+ | * Scan the SHG and optimize the incoupling and crystal position | ||
+ | * Lock the SHG, the voltmeter should display a value around 115 mV (@ 2 W Power Setpoint) | ||
+ | * Activate the powerstabilization | ||
===== Turning off the Laser ===== | ===== Turning off the Laser ===== | ||
* Unlock the laser in DigiLock | * Unlock the laser in DigiLock | ||
Line 28: | Line 30: | ||
* Move the crosshair to the second slope, rightclick on press //PID: Lock to Slop//. | * Move the crosshair to the second slope, rightclick on press //PID: Lock to Slop//. | ||
+ | |||
+ | * **Troubleshooting** | ||
+ | * If signal in DigiLock-Software does not update/ signal is frozen | ||
+ | * If signal in DigiLock-Software does not show the usual error-signal at correct frequency, but is saturated at left and right edge of the frame (see picture) -> Check the ref input for the lock-in-amplifier (near the iodine spectroscopy). Probably the input signal is too low. It should be slightly higher than 2 V (~2.2V works fine).{{ : | ||
===== Lasersystem - Typical power values ===== | ===== Lasersystem - Typical power values ===== | ||
The 285nm SHG was build by Jan Friebe and is described in his diplomathesis. | The 285nm SHG was build by Jan Friebe and is described in his diplomathesis. |