groups:mg:experiment:cameras:wincam [2015/11/30 13:51] – old revision restored (2015/06/03 11:11) ssauergroups:mg:experiment:cameras:wincam [2016/06/10 14:31] (current) klaus
Line 7: Line 7:
 TaperCamD20-15–UCD23:  TaperCamD20-15–UCD23: 
  * WinCamD with 20 x 15 mm 2.27:1 FO taper on the sensor  * WinCamD with 20 x 15 mm 2.27:1 FO taper on the sensor
- * 2/3" CCD sensor for CW Pulsed, 1360 x 1024 pixels, 6.45 um + * 2/3" CCD sensor for CW Pulsed, 1360 x 1024 pixels 
 + * Pixelsize: **ATTENTION** The manual statesit is 6.45 µm. But taking the physical dimension (20x15 cm) and amount of pixels into account, it is rather **14.6772 µm**, which is approximately 2.27 * 6.45
  
 If you have a TaperCamD series camera, press **Alt S** to reach the WinCamD Capture Setup dialog, and enter the appropriate PMF value in the WinCamD capture screen. If you have a TaperCamD series camera, press **Alt S** to reach the WinCamD Capture Setup dialog, and enter the appropriate PMF value in the WinCamD capture screen.
Line 18: Line 18:
  
 ===== Shortcuts ===== ===== Shortcuts =====
 +Capture setup dialog: Alt+S
 Start measurement: Press G Start measurement: Press G
 +Stop measurement: Press S
  
-Stop measurement: Press S+===== Troubleshooting ===== 
 +Sometimes the camera is not getting recognized. Open the capture setup dialog (Alt+S) and unplug + plug the camera. Now, the program should recognize the camera.
  
 ===== Power Limits ===== ===== Power Limits =====