meta data for this page
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
groups:mg:experiment:cameras:wincam [2015/11/30 13:49] – ssauer | groups:mg:experiment:cameras:wincam [2016/06/10 14:31] (current) – klaus | ||
---|---|---|---|
Line 1: | Line 1: | ||
- | ====== | + | ====== |
- | The Camera | + | The Camera |
===== Technical Data ===== | ===== Technical Data ===== | ||
- | https://www.conrad.de/ | + | TaperCamD20-15–UCD23: |
+ | * WinCamD with 20 x 15 mm 2.27:1 FO taper on the sensor | ||
+ | * 2/3" CCD sensor for CW Pulsed, 1360 x 1024 pixels | ||
+ | * Pixelsize: **ATTENTION** The manual states, it is 6.45 µm. But taking the physical dimension (20x15 cm) and amount of pixels into account, it is rather **14.6772 µm**, which is approximately 2.27 * 6.45 | ||
- | ===== Data sheet ===== | + | If you have a TaperCamD series camera, press **Alt S** to reach the WinCamD Capture Setup dialog, and enter the appropriate PMF value in the WinCamD capture screen. |
+ | * TaperCamD20 | ||
+ | Also enter the correct wavelength. | ||
+ | **Congratulations**. You have successfully completed installation. =) | ||
+ | |||
+ | |||
+ | ===== Shortcuts ===== | ||
+ | Capture setup dialog: Alt+S | ||
+ | Start measurement: | ||
+ | Stop measurement: | ||
+ | |||
+ | ===== Troubleshooting ===== | ||
+ | Sometimes the camera is not getting recognized. Open the capture setup dialog (Alt+S) and unplug + plug the camera. Now, the program should recognize the camera. | ||
+ | |||
+ | ===== Power Limits ===== | ||
+ | Power >2.5 x (Beam diam in mm) W, or >10 W total, may damage the ND filter. | ||
+ | The limit is lower for pulsed lasers, especially ns & fs lasers. | ||
+ | |||
+ | | nm | Damage Threshold | ||
+ | | | ||
+ | | 355 | ||
+ | | 630 | ||
+ | | 800 | ||
+ | | 1060 | “ | ||
+ | |||
+ | ===== Data sheet ===== | ||
+ | *WinCamD: {{: |