meta data for this page
  •  

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Next revision
Previous revision
groups:mg:experiment:cameras:wincam [2015/02/19 08:14] – created fimgroups:mg:experiment:cameras:wincam [2016/06/10 14:31] (current) klaus
Line 1: Line 1:
 ====== WinCamD ====== ====== WinCamD ======
  
-The Camera were using belongs to the Chip experiment.+The Camera we are using belongs to the Chip experiment.
  
-====== Technical Data ======+===== Technical Data =====
  
 TaperCamD20-15–UCD23:  TaperCamD20-15–UCD23: 
- WinCamD with 20 x 15 mm 2.27:1 FO taper on the sensor + * WinCamD with 20 x 15 mm 2.27:1 FO taper on the sensor 
- * 2/3" CCD sensor for CW Pulsed, 1360 x 1024 pixels, 6.45 um+ * 2/3" CCD sensor for CW Pulsed, 1360 x 1024 pixels 
 + * Pixelsize: **ATTENTION** The manual statesit is 6.45 µm. But taking the physical dimension (20x15 cm) and amount of pixels into account, it is rather **14.6772 µm**, which is approximately 2.27 * 6.45
  
- +If you have a TaperCamD series camera, press **Alt S** to reach the WinCamD Capture Setup dialog, and enter the appropriate PMF value in the WinCamD capture screen. 
-If you have a TaperCamD series camera, press **Alt S** to reach the WinCamD Capture Setup dialog, and enter the appropriate PMF value in the WinCamD capture screen.  + * TaperCamD20  - 15 - UCD23 Pixel multiply factor = **2.27** 
- * TaperCamD20  - 15 - UCD23 Pixel multiply factor = 2.27+Also enter the correct wavelength.
  
 **Congratulations**. You have successfully completed installation. =) **Congratulations**. You have successfully completed installation. =)
 +
 +
 +===== Shortcuts =====
 +Capture setup dialog: Alt+S
 +Start measurement: Press G
 +Stop measurement: Press S
 +
 +===== Troubleshooting =====
 +Sometimes the camera is not getting recognized. Open the capture setup dialog (Alt+S) and unplug + plug the camera. Now, the program should recognize the camera.
 +
 +===== Power Limits =====
 +Power >2.5 x (Beam diam in mm) W, or >10 W total, may damage the ND filter. 
 +The limit is lower for pulsed lasers, especially ns & fs lasers.
 +
 +| nm    |  Damage Threshold  |  Saturation Irradiance  |
 +|        mW/cm2            |  uW/cm2                 |
 +| 355    Few                ~0.3                   |
 +| 630    >10                ~0.15                  |
 +| 800    “                  ~0.1                   |
 +| 1060  |  “                  ~10                    |
  
 ===== Data sheet ===== ===== Data sheet =====
  
  *WinCamD: {{:groups:mg:experiment:wincamd_man.pdf|}}  *WinCamD: {{:groups:mg:experiment:wincamd_man.pdf|}}