- | ===== Clocklaser ===== | ||
- | ==== Turning on the Laser ==== | ||
- | We use toptica electronics: | ||
- | - Turning the main-key to on. The LED´s from the temperature control will go on! | ||
- | - Push the green button. The green LED from the button will go on! | ||
- | - Switch on the toggle (=Kippschalter) from the laserdiode. The green LED from the laserdiode module will go on! | ||
- | - Switch on the toggle from the TA, if he is seeded! The green LED from the TA module will go on! | ||
- | If the switches are on, then not steps 3. & 4.:-) | ||
- | |||
- | ==== Turning off the Laser ==== | ||
- | Both toggles to switch to off(Laserdiode/ | ||
- | - Push the red button. The green LED will go out! | ||
- | - Turning the main-key to " | ||
- | |||
- | === Turning off the Laserdiode === | ||
- | - Switch off the toggle from the TA. The green LED from the TA module will go out! | ||
- | - Switch off the toggle from the laserdiode. The green LED from the laserdiode module will go out! | ||
- | |||
- | === Turning off or on only the TA === | ||
- | - Switch off or on the toggle from the TA. The green LED from the TA module will go out or on! | ||
- | |||
- | ==== Locking to the resonators ==== | ||
- | |||
- | ==== Littmann design ==== | ||
- | {{ : | ||
- | {{ : | ||
- | |||
- | **Typical efficiencies at 916 nm:** | ||
- | * Isolators: 80 % | ||
- | * AOMs: 50-60 % | ||
- | * Fiber coupling: 50 % | ||
- | |||
- | **Typical values for the laser:** | ||
- | * Frequency: 327.5293 THz (doubled upstairs directly behind SHG: 655.058 566 THz) | ||
- | * Power after ECDL: ~12 mW (@ ~90 mA) | ||
- | * Power after TA: ~200 mW (@ ~1000 mA) | ||
- | * TA current should be <1.2A otherwise the fiber stabilization will soar (=aufschwingen) | ||
- | |||
- | **Typical values in front fibers:** | ||
- | * Power before fiber for wavemeter/ | ||
- | * Power before AtomLabFiber: | ||
- | * Power to R2: ~2.5 mW | ||
- | |||
- | **Typical values for the resonators: | ||
- | * In front of telescope: 2.2 mW | ||
- | * In front of AOM: 1.5 mW | ||
- | * In front of fiber: 450 µW | ||
- | * 100 µW between Isolator and BS | ||
- | * 50 µW in front of Resonator | ||
- | |||
- | **Typical values for locking signals:** | ||
- | * PDH error signal locked ~500 mV PP / unlocked ~200 mV PP | ||
- | * Intensity Error signal ~2 mV PP / DC: | ||
- | * Fiber Stabilization Locked ~200 mV PP / Unlocked ~2 Volt PP | ||
- | |||
- | |||
- | **Manuals and Datasheets** |